OVERVIEW
Designes for the production of pure water and ultra pure water out of tap water
- Aquinity² P versions produce deionised water (< 0.2 µS/cm/ > 5 MOhm x cm) by reverse osmosis and conditioning module in to the external reservoir (35 liter or 70 liter).
- The water can be tapped directly (e.g. for glassware cleaning by hand) or connecting the tank to a laboratory dishwasher.
- Out of the storage tank the ultra pure water is produced and then can be tapped with a flexible dispenser.
- The Aquinity² water purification system is available in different configurations (Reagent, Analytical & LifeScience) to meet the specific requirements of ultra pure water for different applications.
- Analytical is a version with UV lamp for low TOC value. Life Science version comes with UV lamp and Ultrafiltration Module for low TOC and low endotoxin values.
- An optional integrated TOC sensor allows the user to monitor the TOC content permanentely online.
- Besides that the Aquinity² can be configured as bench top or bench-integrated system with external panel and wall mounted (with inclined display).
- The tank can be installed at the side or above the system or even underneath the bench.
- Upon request the tanks can be connected in series to reach a bigger volume of reservoir.

Main technical specifications of the system
- Resistivity/Conductivity
- > 5 MOhm x cm | < 0.2 µS/cm; Type II
- 18.2 MOhm x cm | 0.055 µS/cm; Type I
- Production rate
- max. 10 L/h, optionally 20 L/h
- Total Organic Carbon (TOC)
- < 10 ppb (Reagent)
- < 5 ppb (Life Science)
- < 3 ppb (Analytical)
Flowchart