The LCM Liquid Content Module for the DFA100 measures the moisture of foams with respect to time on different foam height levels. The results are obtained by detecting the foam’s conductivity as a proven method. They provide you with knowledge about a foam’s stability and help you optimize products where moisture plays a key role, such as washing foams.
DFA100LCM
Analyzing liquid content and drainage of foam
Description
SPECIFICATIONS
SPECIFICATIONS
Software
| Software type | ADVANCE |
|---|---|
| Software languages | Chinese (simplified), English, French, German, Japanese, Korean, Portuguese, Russian, Spanish |
| Software modules | Foam analysis (recommended)
Remote Control API (optional) |
Measurement specifications
| Analyzed foam characteristic | foamability and foam stability liquid content and drainage |
|---|---|
| Results | foam height liquid height total height foam capacity maximum foam density expansion rate foam half life time drainage half life time sample temperature liquid content at 7 sensor positions resistance at 7 sensor positions 25 %, 50 % and 75 % liquid content time |
Sample dimensions
| Min. recommended sample volume | 50 mL with 40 mm diameter column 20 mL with 20 mm diameter column |
|---|
Instrument dimensions
| Footprint | 245 mm × 275 mm (W × D) |
|---|---|
| Height | 460 mm |
| Weight (without accessories) | 10 kg |
Environment
| Operating temperature | 15 to 30 °C |
|---|---|
| Humidity | without condensation |
Interfaces
| PC | 1× USB 2.0 |
|---|
Power supply
| Voltage | 100 to 240 V |
|---|---|
| Power consumption | 30 W |
| Frequency | 50 to 60 Hz |
Temperature measurement
| Sensor | PT100 |
|---|---|
| Range | 4 to 90 °C |
| Resolution | 0.1 °C |
| Precision | ±0.1 °C |
| Accuracy | 1/3 DIN B (±0.1 °C at 0 °C / ±0.8 °C at 400 °C) |
| Locations | inside sample liquid |
| External sensor | optional |
Temperature control
| Type of temperature control | optional: double-walled glass column |
|---|---|
| Range of temperature control | 4 to 90 °C (additional thermostat needed) |
Accessories
| Glass columns | 40 mm diameter, temperature control option |
|---|---|
| Filter plates for sparging | diameter: 30 mm |
| Filter plate porosities | G1: nominal maximum pore size: 100 to 160 μm G2: nominal maximum pore size: 40 to 100 μm G3: nominal maximum pore size: 16 to 40 μm G4: nominal maximum pore size: 10 to 16 μm |
| Material of columns and frits | borosilicate glass (norm: ISO 4793) |
| Material of sealings | silicone and FKM |
Camera system
| Diameter of minimum detectable bubble | – |
|---|---|
| Mean field of view size | – |
| Focus | – |
Camera system (standard)
| Connection | – |
|---|---|
| Performance | – |
Illumination
| Types | LED |
|---|---|
| Wave length, dominant (height detection) | 469 nm (optional IR: 850 nm) |
Electrodes
| Material | 35 μm copper, finish: chemical gold |
|---|---|
| Highest sensor position | 185 mm |
| Measured entity | electrical resistance in Ω |
| Theoretical measurement range | 10 Ω to 2 MΩ |
Operating system
| Internal gas flow rate | 0.2 to 1 L/min |
|---|---|
| External gas flow rate | 0.05 to 1 L/min |
| Approved gases | air nitrogen carbon dioxide |
| Approved pressure | 5 ± 0.5 bar |
| Approved temperature | 4 to 90 °C |
| Stirring speed | – |
Line sensor
| Sensor resolution | 1728×1px |
|---|---|
| Spatial resolution | 200 dpi | 0.125 mm |
| Temporal resolution | 20 fps |
| Scanning length | 216 mm |
Shipping & Delivery